BHM Kwiktrak Fixed Gate 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Arjo, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    36636
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0479-2007
  • 사례 시작날짜
    2006-10-23
  • 사례 출판 날짜
    2007-01-11
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2010-01-09
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    patient lift component - Product Code FSA
  • 원인
    The kwiktrak gate stopper may fail and allow the ceiling lift to pass through to the end of the track and fall to the floor.
  • 조치
    Arjo Inc. USA sent recall letters to the direct accounts on 10/30/06, informing them that if the ceiling lift is moved too fast or too forcefully into a closed gate, the gate stopper can be damaged, causing its function to fail. One consequence can be that the gate may remain open, allowing the ceiling lift to pass through to the end of the track. The letters requested the customer to visually inspect their Kwiktrak Gate installation used with their Maxisky-600 Ceiling Lift to determine if there are any signs of impact or burr marks on the closed gate stopper. Illustrations of a damaged and normal gate stopper were included. The customers were requested to stop use of the Kwiktrak track system if any signs of impact or burr marks are found, and contact the Arjo Quality Department at 1-800-323-1245, ext. 6140 for special instructions. If no signs of damage were found, the customers were reminded to make sure that they manually drive the ceiling lift in a gentler manner to prevent damage to the gate stopper, making sure that all personnel using the ceiling lifts are properly trained. The accounts were informed that Arjo personnel will contact them in the next few weeks to install an improved system that will eliminate the risks of failure.

Device

  • 모델명 / 제조번호(시리얼번호)
    Part #700.11550 installed in combination with part #700.15515.
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Michigan, Ohio and Illinois
  • 제품 설명
    BHM Kwiktrak Fixed Gate component of the Kwiktrak rail system used with the Maxi Sky 600 Ceiling Patient Lift; Manufactured by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5. || The fixed gate, part #700.11550, was sold in two kits: a) Part #700.11500 Kwiktrak Gate Kit composed of the fixed gate and a 3-contact interface kit part #700.15515, and || b) Part #700-11505 Double Gate Kit composed of two fixed gates and 3-contact interface kits
  • Manufacturer

Manufacturer

  • 제조사 주소
    Arjo, Inc., 50 Gary Ave Ste A, Roselle IL 60172-1684
  • Source