Biomet Hip System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Zimmer Biomet, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79685
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1537-2018
  • 사례 시작날짜
    2018-04-02
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
  • 원인
    Two lots of different sized modular heads potentially commingled. risks include delay in surgery and dislocation due to use of an incorrect sized head.
  • 조치
    The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 4/2/2018 to its customers. The letter described the product, problem and actions to be taken. Distributors will be notified of the recall via email. Hospital risk managers and distributors with product will be notified via courier. The distributor letter includes instructions to locate and remove the product in their territory, identify hospitals who have previously used the product; return on-hand product to the recalling firm and ensure all products are accounted for. The letter for hospital risk managers instruct the risk managers to assist the recalling firm's sales reps with quarantine of the product; to review notification and ensure affected personnel are aware of the contents; and complete and return the Certificate of Acknowledgement form to CorporateQuality.PostMarket@zimmerbiomet.com, even if you do not have product, and retain a copy. Customers with questions are directed to call customer service at 574-371-3071 between 8AM-5PM EST, Monday through Friday. Alternatively, questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.

Device

  • 모델명 / 제조번호(시리얼번호)
    lot 276760
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    Yes
  • 유통
    Worldwide Distribution: US (nationwide) in states of: AL, AR, FL, GA, IN, KY, LA, MO, MS, OH, OK, TX, and VA; and countries of: Canada, Costa Rica and Chile.
  • 제품 설명
    Modular Head Component, Biomet Hip System; Item No. 163669, Hip Joint, metal/polymer semi-constrained cemented prosthesis || The device is used in treatment of patients who need total hip arthroplasty, and consists of modular components and corresponding instruments.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Zimmer Biomet, Inc., 56 E Bell Dr, Warsaw IN 46582-6989
  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source