U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Vanilmandelic acid test system. - Product Code DCK
원인
Test results are high, outside of control material upper limit specification. results could be as high as 15%.
조치
Cholestech provided its three distributors a notification letter/response form with instructions to distribute the letter to their customers/end users dated August 9, 2006. The letter instructed customers to discontinue use of the lot of cassettes, return remaining cassettes to Cholestech through the Technical Service Representative and to complete and return the response form. The end users returned the signed letters directly to Cholestech so that effectiveness checks could be completed.
Nationwide Distribution --- including states of Virginia, Texas and North Carolina, Ohio, Arkansas, Washington, Florida , California and Michigan.
제품 설명
Cholestech LDX High Sensitivity C-Reactive Protein (hs-CRP) Test Cassette. The Cholestech LDX high sensitivity C-Reactive Protein (hs-CRP) is an in vitro diagnostic test for the quantitative determination of C-reactive protein in whole blood or serum. Measurement of CRP is useful as an aid in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. The test cassettes are for use with the LDX analyzer.
“We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.