CSection Pack 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Windstone Medical Packaging, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    71968
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0159-2017
  • 사례 시작날짜
    2015-04-22
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2016-10-25
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Light, surgical, ceiling mounted - Product Code FSY
  • 원인
    The medtronic cover light handle disposable gloves, part # 13140, of the custom ams kit were recalled because gloves may contain splits or holes compromising sterility.
  • 조치
    Aligned Medical Solutions (AMS) sent the Medtronic Urgent Medical Device for Devon Light Glove letter, dated April 23, 2015 to all affected consignees. Consignees were informed that the sterile kits assembled and distributed by AMS contain the recalled Devon COVER LIGHT HANDLE DISPOSABLE Glove (AMS part #13140 or Medtronic part # 571711) due to splits or holes. Consignees were advised to identify any remaining affected kits and follow the over-labeling instruction. At the time the kit is opened for use, any light gloves should be identified and disposed of pursuant to the medical waste policies in effect at your institution. AMS can provide replacement product if needed or you can contact your local Medtronic sales rep for sterile replacements as well. Due to the manufacturing process and finished sterilized packs in inventory, or at your facility, over labeling of inventory and/or future shipment affected by this recall will be labeled with a sticker attached to each pack. This sticker will read: RECALL NOTICE Medtronic announced the recall l of Devon Light Glove contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall. 1. At the time the kit is opened for use any Light Glove manufactured by Medtronic should be identified and set aside 2. The recalled light glove should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. Consignees should complete the attached Reply Form and fax to AMS at 406-256-9875. For any questions about this recall, please call 321-527-7714. For questions regarding this recall call 404-259-6387.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Numbers/ Expiration Dates:  58207 8/31/2013 61591 1/2/2014 59228 1/19/2014 66522 1/31/2014 61763 3/5/2014 62929 9/10/2014 67485 9/19/2014 67664 9/21/2014 63835 9/22/2014 65290 9/30/2014 56956 6/18/2016 57666 6/23/2016 71060 10/13/2016 68943 10/26/2016 79065 11/9/2017 79842 1/25/2018 80908 2/15/2018
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution to California, Florida, Georgia, Illinois, Kansas, Louisiana, Michigan, New Jersey, Nevada, Texas, Virginia, and Wyoming.
  • 제품 설명
    C-Section Pack, Kit number 002643-10, 002643-9, and AMS4449 || convenience custom kits used for general surgery in hospital operating room