Device Recall 7" (18 cm) Appx 0.24 ml, Smallbore Ext Set w/NanoClave", Clamp, Luer 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 ICU Medical, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    67166
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0688-2014
  • 사례 시작날짜
    2013-12-24
  • 사례 출판 날짜
    2014-01-09
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2014-04-01
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Set, administration, intravascular - Product Code FPA
  • 원인
    Icu medical is recalling certain lots of the nanoclave needlefree connector product line because of a small percentage of devices that may leak.
  • 조치
    icumedical sent an Urgent Medical Device Recall letter dated December 24, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to return completed verification forms. Customers with questions were instructed to contact ICU Customer Service Monday through Friday between the hours of 8:30AM and 4:00PM Pacific Time: (866) 829-9025 and select option 8, or email: customerservice@icumed.com. ICU Medical reported an update on January 6, 2014, an expansion of the recall to additional lots. Updated recall letter was sent out to all customers on additional lots on January 6, 2014 via UPS with tracking. For questions regarding this recall call 949-366-2183.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot No. 2767364
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - USA including UT, TX, GA, WI, VT, NH, AL, OH, TN, IL, CA, KY,HI, NC, and Internationally to CA, BE, IN, KW, and AU.
  • 제품 설명
    7" (18 cm) Appx 0.24 ml, Smallbore Ext Set w/NanoClave", Clamp, Luer || Lock, Part No. A1000, Item No. K7095-001 || The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration set for the administration of fluids to a patient through a cannula placed in the vein or artery.
  • Manufacturer

Manufacturer

  • 제조사 주소
    ICU Medical, Inc., 951 Calle Amanecer, San Clemente CA 92673-6212
  • 제조사 모회사 (2017)
  • Source
    USFDA