Mass spectrometer, clinical use - Product Code DOP
원인
Multiquant md software where under certain conditions a user
can be presented with incorrect quantitative results when using the sum multiple ions feature.
조치
AB SCIEX issued recall notification by letter dated June 16, 2016, to all affected customers. The letter identified the problem, provided temporary measures until a permanent software is available. A response form was to be completed and returned. questions regarding this notice
contact SCI EX at +1 289 982 2712.
Worldwide Distribution - US (Nationwide) Internationally to Canada, Austria, China, Netherlands, Norway, India, Korea, Germany, France, Switzerland, Ireland, Italy, Singapore, Russia, UK, Malaysia,
제품 설명
AB SCIEX QTRAP¿ 4500MD LC/MS/MS System || Mass Spectrometer for In-Vitro Diagnostic Use. || Instrument Part Number (REF): 5031231 || 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.