U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Medical computers and software - Product Code LNX
원인
Ordering salt-based ingredients on an ion-based template resulted in medication errors when calculating amounts of ingredients required.
조치
Baxa Corporation sent a "Safety Alert" letter dated January 19, 2009 via FedEx to all affected customers. The letter described the affected product, problem, and actions to be taken. Customers were instructed to ensure that users create either a salt-based template OR an ion-based template for ordering, and that both methods of ordering are not used in one template. Customers were provided a list of safe practices to follow on the enclosed ISMP Safety Alert. Customers were advised upon receipt of the Safety Alert that Baxa Technical Support will review and address any issues with the customer on their configuration setup for calcium and magnesium. For questions contact Baxa Technical Support at 1-800-678-2292.
Catalogue Numbers: 8300-0045: Abacus Calculator Only (Abacus CE) 8300-0046: Abacus Single Work Station (Abacus SE) 8300-0047: Abacus Multi-Work Station (Abacus ME)