Device Recall Access Free T4 Assay 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Beckman Coulter Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    55196
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2658-2010
  • 사례 시작날짜
    2010-03-24
  • 사례 출판 날짜
    2010-10-01
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-05-07
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Radioimmunoassay, Free Thyroxine - Product Code CEC
  • 원인
    The recall was initiated after beckman coulter determined that kit boxes for access free t4 reagent (ref 33880), lot number 916293 may contain reagent packs labeled as access vitamin b12, (ref 33000), lot 916277. using an access vitamin b12 reagent pack, lot 916277, filled with access free t4 reagents may generate incorrect patient results.
  • 조치
    A Product Corrective Action (PCA) letter was sent on March 24, 2010 to the affected customers. The letter provides the customers with an explanation of the problem identified and instructs them to discard the FT4 kit, Lot 916293 in their stock if the FT4 packs within the kit packaging are labeled as Vitamin B12 Lot 916277. The PCA requested customers: (1) Inspect all Access Vitamin B12 reagent packs that have been removed from their kit boxes, including reagent packs currently loaded on your system and discard any reagent packs labeled as Access Vitamin B12, lot 916277; (2) Inspect your Access Free T4 reagent kit inventory for any reagent packs labeled as Access Vitamin B12, lot 916277 and discard any reagent packs labeled as Access Vitamin B12, lot 916277; (3) Review your historical quality control to ensure that assay performance for Access Vitamin B12 lot 916277 has been within your laboratory specifications. In addition customers were instructed to share this information with their laboratory staff and retain this notification as part of your laboratory Quality System documentation. Complete and return the enclosed response form within 10 days so Beckman Coulter can be assured that you have received this important notification. If customers need assistance or have any questions regarding the notification, they were instructed to contact Customer Technical Support Center at 1-800-854-3633 in the United States and Canada.