Device Recall Accura System for Blood Filtration 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Baxter Healthcare Renal Div 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    33260
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0153-06
  • 사례 시작날짜
    2005-08-29
  • 사례 출판 날짜
    2005-11-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2008-10-29
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Dialyzer, High Permeability With Or Without Sealed Dialysate System - Product Code KDI
  • 원인
    A software anomaly that occurs when the reset totals option is selected during cvvhd therapy with a diaylsate flow rate of less than 600 ml/hour, or cvvh therapy with a predilution flow rate of less than 600 ml/hour, causes balance alarms without an obvious cause.
  • 조치
    Baxter sent Urgent Device Correction letters dated 8/29/05 to all Accura device customers via first class mail. The letter informed the accounts of the balance alarms without an obvious cause as a result of a software anomaly that occurs when the Reset Totals option is selected during either CVVHD therapy with a diaylsate flow rate of less than 600 mL/hour, or CVVH therapy with a predilution flow rate of less than 600 mL/hour. To reduce the probability of a balance alarm occurring or reoccurring for this reason, the customers were instructed to not press the Reset Totals option when the dialysate/predilution flow rates are less than 600 mL/hour. The letter provided the customers with two options to resolve this alarm: 1) With physician approval, increase the flow rate to at least 600 mL/hour briefly (be sure to select and confirm), and then decrease the flow rate back to the original prescribed rate (be sure to select and confirm), and 2) If the flow rate cannot be adjusted as mentioned in option 1, then the current treatment must be ended and a new treatment initiated. Any questions were referred to the Center for One Baxter at 1-800-422-9837, and additional technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt #3, option 4 for Renal Technical Support.

Device

  • 모델명 / 제조번호(시리얼번호)
    All serial numbers
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide, and internationally to China and Belgium
  • 제품 설명
    Accura System for Blood Filtration, product codes 5M5660 and ACCURA01 (international); a hemofiltration system; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085, manufactured in Germany
  • Manufacturer

Manufacturer

  • 제조사 주소
    Baxter Healthcare Renal Div, 1620 Waukegan Rd Bldg R, Mc Gaw Park IL 60085-6730
  • Source
    USFDA