Device Recall ADVIA 2400 Chemistry System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Bayer Healthcare, LLC (Diagnostics Division) 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    35880
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1450-06
  • 사례 시작날짜
    2006-02-17
  • 사례 출판 날짜
    2006-08-29
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-05-24
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Automated Clinical Chemistry Analyzer - Product Code JGS
  • 원인
    Low qc recovery observed on carbamazepine (carb) results immediately following a gentamicin (gent) assay or a digoxin (dig) assay. when gent or dig precedes carb, the carb test result is artificially low (as much as -25%). phny was determined to also exhibit a similar carryover effect on carb. all others were found not to produce the same reagent probe-based carryover effect.
  • 조치
    Bayer HealthCare LLC issued a Support Bulletin to all affected Bayer HealthCare LLC Branches on/about 2/17/06 in order to inform them of the issue and to provide them with instructions for this ''field correction''. Customer Bulletins were also issued to the Branches both in the United States and elsewhere for communication with the affected customers.

Device

  • 모델명 / 제조번호(시리얼번호)
    Part No. 073-A010; all Advia Chemistry System units are serial numbers are involved in this ''field correction'' as well as existing assay lots for CARB, PHNY, GENT and DIG.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide including USA and Puerto Rico.
  • 제품 설명
    Advia 2400 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Bayer Healthcare, LLC (Diagnostics Division), 511 Benedict Avenue, Tarrytown NY 10591-5097
  • Source