Device Recall ADVIA Centaur Calibrator 30 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Siemens Healthcare Diagnostics, Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    78645
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0515-2018
  • 사례 시작날짜
    2017-06-07
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Radioimmunoassay, estradiol - Product Code CHP
  • 원인
    Siemens healthcare diagnostics has confirmed a negative bias for advia centaur enhanced estradiol (ee2) on the advia centaur, advia centaur xp and advia centaur xpt systems when calibrating with calibrator 30 kit lots ending in 21 (c3021) as compared to calibrator 30 kit lots ending in 20 (c3020).
  • 조치
    An Urgent Medical Device Recall letter dated June 7, 2017 was distributed to all customers in the United States and an Urgent Field Safety Notice was distributed to all customers outside the United States who have received ADVIA Centaur Systems Calibrator 30 kit lot ending in 21(C3021). These letters instruct customers who operate the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems to discontinue use of C3021 . The letters inform customers that Calibrator 30 kit lots ending in 20 (C3020) and 23 (C3023) are not impacted and can be used. The letters inform customers who operate ADVIA Centaur CP systems to continue using their inventory of C3021, as they are not impacted by this issue. All affected domestic customers were sent a hard copy of the Urgent Medical Device Recall via Federal Express. Fax-back forms will be used to confirm that the customers received the Urgent Medical Device Recall letter.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Numbers: 00387A21, 16673A21, 36243A21;  00388A21, 15448A21, 51211A21;  20640A21; 51024A21.  Expiration Date: 2018-03-08
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    US and OUS.
  • 제품 설명
    ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. || Siemens Material Number (SMN): 10379810, 10379811, 10697756.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Siemens Healthcare Diagnostics, Inc, 333 Coney St, East Walpole MA 02032-1516