Device Recall Analytical XRay Fluorescence Systems of model family Delta 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Olympus Scientific Solutions Americas 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    74593
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2255-2016
  • 사례 시작날짜
    2016-06-22
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    X-ray fluorescence - Product Code RBY
  • 원인
    Trigger lock option does not always prevent x-ray generation. specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. the correction is via repair by a software update.
  • 조치
    The firm notified customers via an "Important - Electronic Product Radiation Warning" letter. It described the issue; and, told customers to not rely on the Trigger Lock feature until the software upgrade had been completed. The letter also provided instructions for customers to not only determine their software version; but, also instructions to bring their affected device into compliance. The firm will remedy the issue without charge. Customers can contact their local representative or OSSA at ani.technicalsupport@olympus-ossa.com, phone: +1 866 446 6689, fax: +1 781 419 3980, with questions or if additional information is needed.

Device

  • 모델명 / 제조번호(시리얼번호)
    All Models of the Delta XRF Analyzer with Software version 2.5.17.1 and higher
  • 의료기기 등급
    Not Classified
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution.
  • 제품 설명
    Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. || Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium , Delta Professional , and Delta-Pl us.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Olympus Scientific Solutions Americas, 48 Woerd Ave, Waltham MA 02453-3824
  • 제조사 모회사 (2017)
  • Source
    USFDA