Device Recall ASpace SIBD and Arcadius XP L systems 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Aesculap, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    72646
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0422-2016
  • 사례 시작날짜
    2015-11-10
  • 사례 출판 날짜
    2015-12-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2016-09-08
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Screwdriver, surgical - Product Code LRZ
  • 원인
    The flexible shaft of the instruments may break during orthopedic spinal surgery. there is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.
  • 조치
    The firm, Aesculap Implant Systems, sent a "URGENT MEDICAL DEVICE RECALL" letter dated Nov. 23, 2015 to all consignees via certified letter with return receipt. Distributors and sales reps will be notified via e-mail with return receipt. The letter described the product, problem and actions to be taken. The consignees were instructed to review the product recall notification in its entirety and ensure all users your organization and other concerned persons are informed; determine your current inventory; do not destroy any affected product; immediately remove and return the affected instruments-these instruments may be located within Set ST0485; use the U joint screwdriver, ME014R in place of the SJ706R of which is already part of set ST0485; If you require Customer Service assistance please call: Customer Service: 866-229-3002 or Email: SetWorld_Orders.BBMUS_Service@aesculap.com., and complete and return the Product Removal Acknowledgement Form to Aesculap Quality Assurance Department via mail to: Aesculap Inc., 615 Lambert Pointe Drive, Hazelwood, MO 63042; fax to: 610-791-6882 or email to: val.strawn@aesculap.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). A replacement for the SJ607R will be sent to you when one becomes available (ME190R). If you have any questions, contact via email: val.strawn@aesculap.com/Tel: 610-984-9414 or email: Gordon.young@aesculap.com/Tel: 610-984-9043.

Device

  • 모델명 / 제조번호(시리얼번호)
    All Lots
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution: US (nationwide) and country of: Canada.
  • 제품 설명
    Flexible Drill; || SJ723R; NON STERILE; || B Braun Aesculap; || Aesculap AG || These instruments are intended for use during the implantations of the A-Space SIBD and Arcadius XP L systems.
  • Manufacturer

Manufacturer