Device Recall Bard TruGuide Disposable Coaxial Biopsy Needle 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Bard Peripheral Vascular Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    74663
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2547-2016
  • 사례 시작날짜
    2016-07-06
  • 사례 출판 날짜
    2016-08-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2016-11-17
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Biopsy needle - Product Code FCG
  • 원인
    May contain two trocar tip stylets with coaxials or two blunt tip stylets instead of one of each.
  • 조치
    Bard Peripheral Vascular (BVP) sent an Urgent Medical Device Recall Notification letter dated July 6, 2016 to all affected customers via FedEx with proof of delivery notification.. The letter identified the product, the problem, and the action to be taken by the customer. The notification provided customers the following instructions: 1. Do not use or further distribute any affected product. 2. Check all inventory locations within your institution for affected product code I lot number combination listed in the recall notice. If you have further distributed any of the product code I lot numbers, please immediately contact that location, advise them of the recall, forward these instructions and have them return the affected product to BPV. 3. Please remove any identified product from shelves. 4. If used the affected product, complete and return the attached Recall and Effectiveness Check Form indicating no product will be returned. Once the product affected by recall has been removed from inventory; 5. Fill out the Recall and Effectiveness Check Form. Be sure to state the quantities and lot numbers of each recalled product in stock even if no longer have possession of the recalled product. 6. Call the firm's Recall Coordinator Raye Seisinger at 1-800-321-4254 Option #2 Ext 2501 (M-F 6am to 3pm MST) or email at raye.seisinger@crbard.com. Once all information has been verified, the Recall Coordinator will issue a Return Authorization (XC) Number or Consignment Recall Number (XH) to facilitate the expedient return of the product. BPV will provide a credit for your returned product. 7. Fax the Recall and Effectiveness Check Form to BPV at 1-800-994-6772. If cannot FAX the form, please call the BPV recall coordinator at 1-800-321-4254 Option #2 Ext 2501) and report the required information verbally. 8. A mailing label is enclosed to return the affected product. Please mark the outside package as "RECALLED PRODUCT" and i

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Number: REZK0479
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution to the following; CA, LA, VA, IA, AR, SC, PA, MA, IL, MO. Foreign distribution to the following; Canada, Belgium, Australia, Taiwan, China, Argentina.
  • 제품 설명
    Bard TruGuide Disposable Coaxial Biopsy Needle  Mixed Product || Part Number C1816B || The TruGuide Disposable Coaxial Biopsy Needles are intended for use as a guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney, spleen, lymph nodes and various soft tissue lesions.