Device Recall Baxter RenAmin (Amino Acid) Injection 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Baxter Healthcare Corp. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    55016
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1528-2010
  • 사례 시작날짜
    2010-02-15
  • 사례 출판 날짜
    2010-04-29
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-01-29
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Amino Acid Injection - Product Code CGN
  • 원인
    Routine stability testing of the evacuated container product, along with trending of results, indicates that the ph will exceed the upper specification limit prior to the expiration date of this product. baxter decided to remove all potentially affected products even though there have been no out-of-specification ph results at the 0, 6 and 12 month intervals. the recall was initiated on 2/15/10.
  • 조치
    Urgent Product Recall letters dated 2/15/10 were mailed on the same date via first class mail to the affected direct accounts to the attention of the Director of Pharmacy, the Director of Materials Management, and the Director of Risk Management. The accounts were informed of the potential for the pH of the products to exceed the upper specification limits for pH prior to the product expiration date. The accounts were requested discontinue use and segregate the affected lots of products from inventory and call Baxter Healthcare Center for Service at 1-888-229-0001, Monday through Friday, 7:00 AM to 6:00 PM CT, to arrange for the return of the affected product for credit. The accounts were also requested to complete and fax back to Baxter the enclosed reply sheet confirming receipt of the letter and the amount of each product being returned. If the accounts further distributed the products, they were requested to forward the communication to those accounts. Questions regarding the communication are directed to the Medical Information Services at Baxter at 1-800-933-0303.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot number: G076802
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    United States -- Nationwide and Puerto Rico, Canada and Saudi Arabia
  • 제품 설명
    Baxter RenAmin (Amino Acid) Injection; Pharmacy Bulk Package Not for Direct Infusion, Rx Only, 250 mL type II glass bottles; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA, Made in USA; product code 2A6222.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073
  • Source
    USFDA