U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Knee Prosthesis - Product Code JWH
원인
Mislabeled: the packaging of the two products was mixed up, resulting in the products being labeled with an incorrect size. if implanted, the product may not fit properly, which could require repeat surgery. the problem could also result in a delay in the procedure.
조치
Distributors were notified via recall letter (Urgent Medical Device Removal Notice) dated 2/12/08 to locate the devices in their possession and at hospital sites, discontinue use, and return them to Biomet. Hospitals having received the devices will be provided with a recall letter dated 2/12/08 which describes the nature of the problem and the removal of these instruments. A third letter dated 2/12/08 was provided for implanting physicians.