Device Recall BlomSinger Adjustable Tracheostoma Valve II 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Helix Medical LLC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    66423
  • 사례 위험등급
    Class 3
  • 사례 연번
    Z-0128-2014
  • 사례 시작날짜
    2013-10-01
  • 사례 출판 날짜
    2013-11-01
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2014-01-21
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Prosthesis, laryngeal (taub) - Product Code EWL
  • 원인
    Inhealth technologies (iht) is recalling the blom-singer adjustable tracheostoma valve ii (atsv ii) because the body component of the device is missing from the packaging. the atsv ii valve product cannot be used as intended without the product body.
  • 조치
    A phone script and a customer notification letter are provided. The recall was initiated on 10/1/13 and customers will be notified that InHealth Technologies, a division of Helix Medical, LLC has initiated a recall of the following 3 lots of the Blom-Singer Adjustable Tracheostoma Valve (BE8025H and BE8025H-R3): 1057760, 1057919 and 1059222. The phone script provides the problem identified and the actions to be taken. Customers are instructed to complete the Customer response form and return by email to belinda.jackson@helixmedical.com, fax to (805) 566-5381 or by mail to InHealth Technologies, 1110 Mark Avenue, Carpinteria, CA 93013. Customers are also instructed to Contact UPS at 1-800-742-5877 and provide In Health Technologies UPS account no., 846718. UPS will come to your site, package the recalled product(s) and ship it to InHealth Technologies at InHealth Technologies expense. Customers are instructed to respond within 10 days.

Device

  • 모델명 / 제조번호(시리얼번호)
    LOTS: 1057760, 1057919 and 1059222
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Croatia, Germany, Ireland, and Italy.
  • 제품 설명
    Blom-Singer Adjustable Tracheostoma Valve II || Models: BE8025H and BE8025H-R3
  • Manufacturer

Manufacturer