U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Guide Catheter - Product Code DQY
원인
Gaps were identified between polymer segments on the 7f wiseguide guide catheter in which the ptfe was not properly laminated to the braid of the catheter. this delamination of the ptfe from the braid could compromise the device delivery.
조치
Affected customers were mailed an Urgent Medical Device Recall or Correction - Immediate Action Required letter, dated 05/22/2006. The letter describes the issue, states to immediately discontinue use of and segregate recalled product, provides instructions to ship product back to Boston Scientific and asks for the Reply Verification Tracking Form to be faxed to Boston Scientific in acknowledgement.
Catalog number 19500-448, Lot # 655713; Catalog number 19500-498, Lot # 655721; Catalog number 19500-08, Lot # 657310; Catalog number 19500-09, Lot # 657311; Catalog number 19500-132, Lot # 655680; Catalog number 19500-27, Lot # 655695;
Nationwide- States of AZ, CA, CO, FL, HI, KS, MI, NV, OK, TX, VA.
제품 설명
Boston Scientific Wiseguide Guide Catheter 7F, .076 '', Femoral Approach, Manufactured for Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN USA 55311. Nonpyrogenic.