Device Recall BuckyDiagnost with Optimus Generator 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Philips Healthcare Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    58077
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0905-2011
  • 사례 시작날짜
    2011-01-17
  • 사례 출판 날짜
    2011-03-04
  • 사례 상황
    Completed
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Diagnostic X-Ray High Voltage Generator - Product Code IZO
  • 원인
    Affected systems were determined not to prevent run-on of a selected technique factor if the technique selector and the exposure control were activated simultaneously.
  • 조치
    A Customer Information BuckyDiagnost Diagnostic X-Ray System letter, dated January 6, 2011, was sent to customers. The letter identifies the affected product, the reason for the recall, actions that are to be taken by customers, and actions that are planned by the firm. Customers are advised to make sure that the desired exposure values are displayed and that they have released the technique button BEFORE they start the exposure by pressing the exposure switch whenever manually changing the exposure parameters and in order to avoid the problem. Philips will issue a Field Change Order (FCO) to resolve the issue by updating the system software free of charge. Philips will contact customers to implement the FCO. Further information or support concerning this issue can be obtained by contacting your local Philips representative. The Field Service Engineers began software implementation the week of March 21, 2011.

Device

  • 모델명 / 제조번호(시리얼번호)
    704031 / 70408
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution
  • 제품 설명
    BuckyDiagnost general purpose x-ray system with Optimus Generator. || General Purpose Diagnostic Radiography.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Philips Healthcare Inc., 3000 Minuteman Road, Andover MA 01810
  • 제조사 모회사 (2017)
  • Source
    USFDA