Device Recall Capintec Captus 3000 Thyroid Uptake System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Capintec Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    62982
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2312-2012
  • 사례 시작날짜
    2012-07-24
  • 사례 출판 날짜
    2012-09-05
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2013-12-10
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Probe, uptake, nuclear - Product Code IZD
  • 원인
    Customers who received a capintec captus 3000 thyroid uptake system with software revisions 1.22, 1.23, 1.24, 1.25, and 1.26 may encounter an error while using the thyroid uptake protocol ?measure liquid and subtract residual." this protocol was a new feature added to software revision 1.22 in january 2011. the error occurs only if the user does not measure the residual dose immediately after me.
  • 조치
    The firm, CAPINTEC, INC., sent a "SOFTWARE ERROR ALERT" letter dated July 20, 2012 to all affected consignees/customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the four measuring steps outlined in the letter; DO NOT PRINT, EXIT THE THYROID UPTAKE MODULE, OR MEASURE ANOTHER PATIENT BEFORE YOU HAVE COMPLETED THESE FOUR MEASUREMENT STEPS; and follow the attached addendum for step-by step instructions. Capintec will send corrected software to all affected customers in approximately 2-3 weeks. The customers are instructed to contact Capintec Customer Support Representative if you have any questions or concerns regarding this notification, or if you do not receive your replacement software. Capintec Customer Support can be reached by phone at 1-800-631-3826 or by email at getinfo@capintec.com or by visiting our website at www.capintec.com.

Device

  • 모델명 / 제조번호(시리얼번호)
    Catalog numbers for new systems: 5430-0076, 5430-0077, 5430-0103, and 5430-0087. Catalog numbers for conversion systems: 5430-0118 and 5430-0119. Serial numbers for new systems: 901006 through 901394, inclusive. Serial numbers for conversion systems: CNV 674 through CNV 730, inclusive. Also affected are devices which were serviced and devices for which customers received software upgrades between January 20, 2011 and June 30, 2012 with Revision 1.22, 1.23, 1.24, 1.25, or 1.26. (NOTE: Item number 5430-0075, Captus 3000 well counter system, is excluded. This system does not have the ability to perform thyroid uptake procedures.)
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide distribution: USA (nationwide) and countries of: Argentina, Belgium, Brazil, Canada, China, Greece, Hong Kong, India, Israel, Malaysia, Philippines, Spain, South Korea, Thailand, and Turkey.
  • 제품 설명
    Capintec Captus 3000 Thyroid Uptake System || Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform thyroid uptake procedures, which consist of the measurement of organ uptake and circulating radioactivity.
  • Manufacturer

Manufacturer