Device Recall Chemetron 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Allied Healthcare Products Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    52383
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2031-2009
  • 사례 시작날짜
    2009-05-15
  • 사례 출판 날짜
    2009-09-10
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-04-23
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    compressed gas cylinder and valve - Product Code ECX
  • 원인
    Medical valves do not meet quality specifications which may cause the valve to fail and result in uncontrolled release of oxygen when the valve is in the open position.
  • 조치
    The recalling firm first notified their customers via a letter dated 5/15/09 issued regular mail explaining the reason for recall and requesting the cylinders with specific post valves be returned. A second letter dated 6/1/09 was issued via regular mail which provided specific shipping dates that each customer received the suspect product. Neither of these two letters requested subrecall. A third letter dated 6/17/09, which included copies of the first two letters, via certified mail. This letter informed the customer the recall was to be conducted to the user level and requested their customer conduct a subrecall. Requested to remove the cylinder from service and call Allied Healthcare Products for return and a replacement at 1-800-268-1661.

Device

  • 모델명 / 제조번호(시리얼번호)
    InterMed 08-05-3B and InterMed 09-01-1B located on the post valve.
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Class II Recall Worldwide Distribution including states of NY, NC, KY, PA, WI, NH, NJ, CA, CT, FL and Singapore.
  • 제품 설명
    First Responder Kit containing a 240-liter aluminum oxygen cylinder with individually packaged regulator, cannula, and bag mask resuscitator, all packaged in a shipping container labeled as: Product #L903, Desc FIRST RESPONDER KIT, Catalog #L903. The brand and firm name on the label affixed to the cylinder is Chemetron, Allied Healthcare Products Inc., St. Louis, MO.
  • Manufacturer

Manufacturer