Device Recall CIOS ALPHA 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Siemens Medical Solutions USA, Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    73546
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1281-2016
  • 사례 시작날짜
    2016-02-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-11-08
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Interventional fluoroscopic x-ray system - Product Code OWB
  • 원인
    Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. the risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.
  • 조치
    Siemens sent a Customer Safety Advisory Notice dated February 18, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers that Siemens is planning to replace the affected parts as a precautionary measure. A Siemens service engineer will contact each customer to schedule implementation of this precautionary measure.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Number: 10308191
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    US Nationwide Distributed in the states of AR, CA, NE and UT
  • 제품 설명
    CIOS ALPHA (Image-intensified fluoroscope x-ray system) || Product Usage: || The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Siemens Medical Solutions USA, Inc, 40 Liberty Blvd, Malvern PA 19355-1418
  • 제조사 모회사 (2017)
  • Source
    USFDA