Device Recall Colleague CX Infusion Pump 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Baxter Healthcare Corp. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    59270
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2853-2011
  • 사례 시작날짜
    2011-07-05
  • 사례 출판 날짜
    2011-07-19
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-01-20
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Pump, infusion - Product Code FRN
  • 원인
    While replacing a defective pump head mechanism and performing calibration, baxter's service center may have failed to apply glyptol (a red insulating varnish) on a potentiometer that helps maintain pump accuracy for detecting an air bolus.
  • 조치
    Baxter Healthcare Corporation telephoned its customers on July 1, 2011 and sent a follow up "URGENT PRODUCT RECALL" letter dated July 6, 2011, informing them that the Glyptol (a red insulating varnish) may not have been applied on the Air-In-Line (AIL) circuit board. The letter described the product, problem and actions to be taken. The customers were instructed to locate the affected pumps, remove from service and return to Baxter for inspection as soon as possible. If GLYPTOL is found to be missing, recalibration will be performed and GLYPTOL will be applied per requirement before returning the pumps back to the customer at no cost. The customer was instructed to complete and sign the attached customer reply form. The letter states that if the COLLEAGUE Pumps were distributed to other facilities, please forward communication to the pump owner as appropriate. For any questions contact The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm Central Time, Monday through Friday.

Device

  • 모델명 / 제조번호(시리얼번호)
    Product Code: 2M8161, Lot Numbers 13060235CC and 13090748CC
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide (USA) Distribution -- including the state of Pennsylvania.
  • 제품 설명
    Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division. Product Code: 2M8161. || Product Usage: Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients. Electronic infusion pumps indicated for continuous or intermittent delivery of solution through clinically acceptable routes of administration, such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of fluid spaces applications.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073
  • Source
    USFDA