Device Recall Colleague CX Volumetric Infusion Pump 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Baxter Healthcare Corp. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    57515
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0931-2011
  • 사례 시작날짜
    2010-12-01
  • 사례 출판 날짜
    2011-01-21
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2011-05-13
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Pump, infusion - Product Code FRN
  • 원인
    While the infusion pump was being serviced at the baxter service center, it was inadvertently routed to the packaging area before the final inspection.
  • 조치
    Baxter telephoned the one customer on December 1, 2010, informing them that that an outgoing Quality Control Evaluation was not performed on their Colleague pump serial number 15030799CC during the last service. Since this was not conducted on this pump, the device may not function as designed and could lead to an interruption of therapy. Baxter requested that they locate the pump and remove it from use as soon as possible. Baxter will then send a shipping container for retrieval, reinspection and return of the device back to the customer at no cost. For questions regarding this recall call (800) 437-5176.

Device

  • 모델명 / 제조번호(시리얼번호)
    Product Code: 2M8161, serial number 15030799CC
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Texas
  • 제품 설명
    Baxter Colleague Single Channel Volumetric Infusion Pump. Baxter Healthcare Corporation, Medication Delivery Division. Product Code: 2M8161
  • Manufacturer

Manufacturer

  • 제조사 주소
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073