Unit, cryosurgical, accessories - Product Code GEH
원인
An increase in complaint trending for a loss of audible sound associated with the timer has been observed.
조치
The firm, CSA Medical, sent a letter to its consignees 11/14/2014, via Fedex overnight . The letter explains the event and asks consignees to us a verbal call of time during the cryogen spray period until the root cause is determined and a permanent correction is implemented. Consignees are also asked to complete an acknowledgment form and return it to CSA Medical via, fax, email, telephone, etc.
US Distribution to states of: AR, CA, CO, CT, DC, FL, GA, IL, IN, LA, MA, MD, MI, NC, NJ, NY, OK, PA, RI, SC, TN, TX, UT, VA, and WV.
제품 설명
CSA Medical truFreeze System; Model: CC3-01, a cryogenic surgical device || ***The involved units are those with the newest version of a panel PC referred to as POC-127. These panels are identified by serial number. *** || The truFreeze is intended for cryogenic destruction of tissue requiring either active or passive venting during surgical procedures. The truFreeze is indicated for use as a cryosurgical tool in the fields of dermatology, gynecology, and general surgery, to ablate benign and malignant lesions.