Gas-Machine, Anesthesia - Product Code BSZ
원인
Anesthetic agents could be delivered at concentrations above the concentration set on the s/5 and as/3 anesthesia delivery units (adu) having serial numbers within a specified range.
조치
Recall letters dated February 03, 2004 stated that a firm representative would contact the consignee to schedule a visit to replace software and hardware to correct the problem.
The product was distributed nationwide to hospitals in the Unites States.
제품 설명
Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units || (The S/5 units are a later version of the AS/3 units. The conversion from AS3 to S5 occurred during the summer of 2000. This change was mainly a product name change, with a slight color variation to the shade of white being used externally on the machine. There were no design or feature and function modifications when the name change occurred.)