Device Recall Drill Bit QuickConnect 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Zimmer Biomet, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    77417
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-3193-2017
  • 사례 시작날짜
    2017-05-22
  • 사례 출판 날짜
    2017-09-22
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Instrument, surgical, orthopedic, ac-powered motor and accessory/attachment - Product Code HWE
  • 원인
    A raw material anomaly was discovered during inspection at zimmer biomet, and an investigation by the supplier determined that 4 lots of raw material could have similar anomalies. the anomaly has the potential to be on or below the surface and can increase the risk of instrument fracture. accordingly, all products manufactured with the affected raw material are being removed from the field.
  • 조치
    On 5/22/2017, Zimmer Biomet distributed Urgent Medical Device Recall notices to their customers and distributors. On 12/19/2017, Zimmer Biomet sent Urgent Medical Device Recall notices to their customers due to a recall expansion involving one (1) lot for product # 20. Customers are advised to review the notice and ensure that affected personnel are aware of the contents. All stock should be inspected and product quarantined. A Zimmer Biomet sales representative will remove the affected product from your facility. Customers should complete and return the Certificate of Acknowledgement via email to: corporatequality.postmarket@zimmerbiomet.com and retain the copy for your files. Customers with questions may call the customer call center at 574-371-3071, Monday through Friday 8 am - 5 pm, EST. *For Distributors Your Responsibilities include the following: 1. Review this notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1  Inventory Return Certification Form. a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com within three (3) days. 4. Immediately return all affected product from your distributorship and affected hospitals within your territory along with a completed Attachment 1  Inventory Return Certification Form to Zimmer Biomet. a. For each return, send a copy of the completed acknowledgement form via email to: corporatequality.postmarket@zimmerbiomet.com. b. Include a hardcopy of Attachment 1 with your shipment for immediate processing. c. Include a copy of Attachment 3  Certificate of Sterilization with returned instruments. d. Mark the outside of the returns box(es) clearly with FIELD ACTION. 5. Note that any hospitals that received direct shipments of this product from Zimmer Biomet will be sent a copy of the Risk Manager Field Action Notice direc

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot Numbers: 63545255, 63562124, 63566199, 63566218, 63584030, 63398778, 63426469, 63431693, 63431763, 63444579, 63444581, 63444582, 63444584, 63444587, 63451645, 63451647, 63451648,  63451649, 63451651, 63451655, 63462151, 63462166, 63462170, 63462174, 63474142, 63478948, 63484963, 63484966, 63484973, 63492362, 63492363, 63505347, 63505348, 63510772, 63510773 &  63524431.
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
  • 제품 설명
    Drill Bit Quick-Connect 2.0 mm Diameter 100 mm Length, Item Number/EDI 00480610020, Nonsterile
  • Manufacturer

Manufacturer

  • 제조사 주소
    Zimmer Biomet, Inc., 1800 W Center St, Warsaw IN 46580-2304
  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    USFDA