Device Recall EDWARDS AQUARIUS HEMOFILTRATION SYSTEM 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Edwards Lifesciences, Llc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    53554
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-0653-2010
  • 사례 시작날짜
    2009-03-16
  • 사례 출판 날짜
    2010-01-27
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-01-18
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Dialyzer, high permeability with or without sealed dialysate system - Product Code KDI
  • 원인
    There have been reports of clinically significant fluid imbalance when a certain level of fluid imbalance is detected the aquarius will trigger an alarm, however, users are able to override this alarm and continue therapy. it is possible to remove too much fluid from or replace too much fluid to the patient. in extreme cases, this may result in serious injuries or death.
  • 조치
    Edwards is continuing a field corrective action related to the possibility of fluid imbalance when using the Aquarius. In order to mitigate some of the risk associated with this problem, Edwards sent notifications to consignees advising of mitigation steps. Consignees were provided: a description of the affected product, problem, and potential hazard; advice on action to be taken; and revised device labeling. Additionally, consignees were notified of a planned software upgrade to prevent users from bypassing the fluid balance alarm more than five times in a 20-minute period. Although Edwards does not believe there is a need to stop using the Aquarius system at this time, clinicians should weigh the risks/benefits to the patients before continuing the use of the device Edwards first provided some of this information to consignees in a Field Safety Notice on March 16, 2009. On January 11, 2010, Edwards provided an Urgent Product Recall letter to consignees which updated and expanded the earlier communication. The FDA classified this as a Class I recall.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model GEF09800 all units/serial numbers.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution.
  • 제품 설명
    Edwards Lifesciences Aquarius System Model GEF09800.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Edwards Lifesciences, Llc, 1 Edwards Way, Irvine CA 92614-5688