Device Recall ENCORE(R) UNDERGLOVE 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Ansell Healthcare Products LLC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    59155
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2843-2011
  • 사례 시작날짜
    2011-06-10
  • 사례 출판 날짜
    2011-07-18
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2013-01-10
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Powder free latex surgeons gloves - Product Code NPD
  • 원인
    Gloves do not meet the tensile requirements of astm d 3577 reducing the shelf life of the product from 5 years to 2 years.
  • 조치
    Ansell Healthcare Products LLC sent an Urgent Voluntary Recall letter dated June 9, 2011, to all affected customers on June 10 and 13, 2011. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disconstinue use and distribution of any of the affected product and return them to the firm. If the customer distributed gloves further, customers were advised to forward a copy of the notification letter along with their tracking number to their customers. Customers were instructed to complete and return the attached Response Form to Ansell by fax at 1-800-722-8155. Returned product would be replaced or credited to the customer.

Device

  • 모델명 / 제조번호(시리얼번호)
    All codes distributed from February 2008 through May 2011
  • 유통
    Nationwide Distribution
  • 제품 설명
    ENCORE(R) UNDERGLOVE, Latex, Powder-Free, Smooth Surgical Gloves, REF 2018490, Size 9, Sterile, Manufactured for Ansell Healthcare, 1635 Industrial Road, Dothan, AL 36303, Made in Sri Lanka || Surgical gloves
  • Manufacturer

Manufacturer

  • 제조사 주소
    Ansell Healthcare Products LLC, 1635 Industrial Rd, Dothan AL 36303-5750
  • 제조사 모회사 (2017)
  • Source
    USFDA