Device Recall Endopath Xcel with Optiview Technology 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Ethicon Endo-Surgery Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79801
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1886-2018
  • 사례 시작날짜
    2018-03-26
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Laparoscope, general & plastic surgery - Product Code GCJ
  • 원인
    Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. in addition, insufficient lubrication could result in a remote chance of introducing a minute particle if repeated insertions of the obturator or laparoscopic device damage the seal.
  • 조치
    The firm notified customers and distributors by letter on March 26, 2018, and March 27, 2018, respectively. Customers were informed of the issue and potential risk of prolonged surgery or introduction of minute particles. Customers were instructed to: 1. Examine your inventory immediately to determine if you have any products subject to this recall on hand and quarantine such product(s). 2. Remove the products subject to this recall from your inventory and communicate the issue to all relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any of the devices subject to this recall have been forwarded to another facility, please contact that facility to arrange return. 3. Complete the Business Reply Form confirming receipt of this notice and fax or email it to Stericycle at 1-888-965-5794 or Ethicon3546@stericycle.com within three (3) business days. Please return the BRF even if you do not have the product lot subject to this recall. 4. Customers are required to return all unused devices subject to this recall that are in their inventory immediately. 5. To return unused devices subject to this recall photocopy the completed BRF, place it in the box with the subject product(s), and affix the pre-paid authorized shipping label included with this recall notification letter. Distributors were instructed to: 1. Examine your inventory immediately to determine if you have the product lot subject to this recall on hand and quarantine such product(s). 2. Remove the product lot subject to this recall and communicate the issue to all relevant materials management personnel, or anyone else in your facility who needs to be informed. 3. Complete the Business Reply Form confirming receipt of this notice and fax or email it to Stericycle at 1-888-965-5794 or Ethicon3546@stericycle.com within three (3) business days. Please return the BRF even if you do not have the product lot subject to this rec

Device

  • 모델명 / 제조번호(시리얼번호)
    Lots P4T745 P4T88Y P4T906 P4T91H P4TA60 P4TC8M P4TD0G P4TD2K R40015 R4011E R4025E R4035E R4042Z R4047X R4065E
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide domestic distribution, including PR. International distribution to Saudi Arabia, United Arab Emirates, Belgium, Singapore, Taiwan, South Korea, Canada, Chile, Mexico, Japan, and Argentina.
  • 제품 설명
    5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm, 75mm shaft length, 2B5ST
  • Manufacturer

Manufacturer

  • 제조사 주소
    Ethicon Endo-Surgery Inc, 4545 Creek Rd, Blue Ash OH 45242-2803
  • 제조사 모회사 (2017)
  • Source
    USFDA