Thermal Infusion Fluid Warmer - Product Code LGZ
원인
1. fluid contamination of the electronic circuitry can cause short circuits resulting in permanent fuse failure or localized circuit board overheating with possible smoldering and discoloration.
2. electromagnetic (emi) interference associated with the warmer may produce artifact in ecg, eeg, or emg records (cardiac or neuro monitoring).
조치
Vital Signs, Inc. issued Product Safety Notification on March 23, 2009 informing users of potential EMI and fluid ingress issues. All warmers in the field will be replaced beginning in May 2009.
Worldwide Distribution -- USA, Philippines, Saudi Arabia, India, Israel, Australia, Mexico, Germany, Republic of Singapore, China, Guatemala, Switzerland, Dubai United Arab Emirates, New Zealand, Malaysia, India, Ecuador, & United Kingdom.
제품 설명
Model 100 enFlow Fluid/Blood Warmer || Model Numbers: 980100, 91000103, 91000153, 91000154, 91000156 || The device is a medical point-of-care device designed to warm blood and IV fluids.