Device Recall Enzyme Immunoassay, Vancomycin 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Roche Diagnostics Operations, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    74312
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2304-2016
  • 사례 시작날짜
    2016-05-26
  • 사례 출판 날짜
    2016-08-01
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-03-23
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Radioimmunoassay, vancomycin - Product Code LEH
  • 원인
    Roche diagnostics operations, inc. has issued a voluntary recall for the online tdm vancomycin assay on the cobas c 311/501/502 analyzers and the modular analytics p module due method sheets that state an incorrect method comparison against the cobas integra 800 analyzer.
  • 조치
    Roche sent an Urgent Medical Device Correction letter dated May 26, 2016, to all affected customers via UPS Ground (receipt required) . The letter identified the product the problem and the action needed to be taken by the customer. Customers will be informed to disregard the method comparison data in the package insert and to follow the instructions outlined in UMDC 16-083. Non-responding accounts will be defined as users for which confirmation of receipt of the Urgent Medical Device Correction cannot be obtained. These accounts will be monitored on an ongoing basis and follow-up attempts will be made accordingly. Affected product under Roche control was placed on hold. Affected product will be reworked to add a sticker referring to customers to UMDC 16-083. Roche is not requesting return of the affected product. Any returned product will be managed according to current procedures. The incorrect method comparison will be removed from the method sheets. Updated method sheets will be released at a later time. Customers may contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336.

Device

  • 모델명 / 제조번호(시리얼번호)
    04491050190  ONLINE TDM Vancomycin 100 tests 05108420190  ONLINE TDM Vancomycin 200 tests 04642490190  Hitachi Vancomycin (Modular P) 04642481190  Hitachi Vancomycin (917, MOD P)
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN,TX, UT, VA, VT, WA, WI, WV, WY Foreign:None VA/DOD: See list below
  • 제품 설명
    ONLINE TDM Vancomycin 100 tests; ONLINE TDM Vancomycin 200 tests; Hitachi Vancomycin (Modular P); Hitachi Vancomycin (917, MOD P) || The ONLINE TDM Vancomycin assay is for the quantitative determination of Vancomycin in human serum or plasma on Roche automated clinical chemistry analyzers.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Roche Diagnostics Operations, Inc., 9115 Hague Rd, Indianapolis IN 46256-1025
  • 제조사 모회사 (2017)
  • Source
    USFDA