Device Recall EOS System X ray beam 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 EOS Imaging 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    66833
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0529-2014
  • 사례 시작날짜
    2013-11-07
  • 사례 출판 날짜
    2014-01-09
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2015-01-27
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Solid state x-ray imager (flat panel/digital imager) - Product Code MQB
  • 원인
    Eos imaging discovered during production internal testing that the x-ray beam collimation set up by the operator at the user interface of the acquisition workstation is different from the actual collimation of the x-ray beam operated by the eos system during the x-ray acquisition.
  • 조치
    CDRH approves the following conditions for EOS Imaging SA: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator notes below what are to be included in the notification. The draft letter provided on November 28, 2013, is deficient for the following reasons: a. It does not provide a statement that all repairs will be made without charge (21 CFR 1003.21(3)). b. It does not provide instructions, with respect to the use of the product pending the correction of the defect (21 CFR 1003.21(1). Specifically, the letter should include directions for the user to determine if the collimation defect is currently present in their system, which would initiate immediate correction, rather than at the next semi-annual service. 2. The customer notification letter will be sent in the manner required by 21 CFR 1003.21(b). 3. The repairs outlined are to be completed by July 1, 2014. All systems under service contracts (which includes semi-annual service) would be visited between receipt of this letter and July 1, 2014. All repairs, including those to systems which are not under service contracts, are also expected to be completed by July 1, 2014. EOS may proceed with implementation of the CAP (subject to the conditions noted above. The customer notification letter may be sent out immediately upon addressing the above deficiencies. However, it is suggested that you provide a copy of the revised letter to FDA for approval prior to issuance. For further questions please call (678) 564-5400.

Device

  • 모델명 / 제조번호(시리얼번호)
    EOS System
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution
  • 제품 설명
    EOS System X- ray beam || Digital radiography system used in general radiographic examinations.
  • Manufacturer

Manufacturer

  • 제조사 주소
    EOS Imaging, 10 rue Mercoeur 4 Ieme Etage, Paris France France
  • 제조사 모회사 (2017)
  • Source
    USFDA