Device Recall Epic VR Model V196 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 St Jude Medical CRMD 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    46870
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1325-2008
  • 사례 시작날짜
    2008-01-16
  • 사례 출판 날짜
    2008-08-19
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-04-03
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
  • 원인
    A condition that could lead to a ventricular sensing anomaly in epic and atlas family of lmplantable cardioverter defibrillators (icds) has been identified. a loss of ventricular sensing would prevent the icd from being able to detect an arrhythmia. if an arrhythmia occurred and the device was unable to detect it, then a patient would not receive potentially life saving therapy.
  • 조치
    A January 16, 2008, letter entitled, "Important Device Upgrade Information" was sent to physicians by FedEx. In the US, sales representatives will follow-up with the physician and will upgrade all programmers. For the rest of the world, it will be monitored by various St. Jude offices and local distributors with this information being fed back to St. Jude Medical CRMD in Sylmar. In addition, St. Jude Medical issued a press release on January 18, 2008 and information concerning this advisory can be found on St. Jude's website, www.sirn.com.

Device

  • 모델명 / 제조번호(시리얼번호)
    all serial numbers
  • 유통
    Nationwide and Internationally: Argentina, Australia, Austria, Barbados, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Hong Kong, Hungary, India, Italy, Japan, Korea, Malaysia,Mexico, Netherlands, New Zealand, Norway, Peru. Poland, Portugal, Republic of Panama, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad, United Kingdom, Uruguay and Venezuela.
  • 제품 설명
    Implantable Cardioverter Defibrillator || EPIC + VR Tiered-therapy cardioverter/defibrillator, Model V-196
  • Manufacturer

Manufacturer

  • 제조사 주소
    St Jude Medical CRMD, 15900 Valley View Ct, Sylmar CA 91342-3577
  • Source
    USFDA