Device Recall EXACTECH 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Exactech, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    77004
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2665-2017
  • 사례 시작날짜
    2017-05-25
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Rongeur - Product Code HTX
  • 원인
    Instructions for use (ifu) is updated to clarify the proper technique and instrumentation to ream the glenoid, i.E., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.
  • 조치
    On May 22, 2017 Exactech sent a Notification to all their Direct Accounts an Distributors telling them that a new updated technical techniques LIT # 718-01-30 Rev J is now available in their website (www.exac.com) or can be requested by contacting the Customer Service Representative at 800-392-2832, fax 352-378-2617 or via e-mail at Kaya@exac.com. This updated the technical techniques for implantation for Equinoxe Platform Shoulder System to clarify the proper technique for use of the Pilot Tip Reamers, i.e., to include caution statement that the surgeon should avoid applying a bending force to the pilot tip reamer and should avoid using the reamer to retract the humeral head, as this may result in instrument breakage. The notification also request to forward and informed this notice to all the accounts that have received this product and to complete and return the attached Field Advisory Notice Respond Form within 5 days of receipt. For more information please contact: Kaya Davis, Inventory Representative at kaya@exac.com or by phone at (352) 377-1140

Device

  • 모델명 / 제조번호(시리얼번호)
    All Lots
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands and United Kingdom
  • 제품 설명
    EXACTECH Equinoxe Narrow Modular Reverse Pilot-Tip Reamer // Catalog #: 321-45-38, Catalog #: 321-45-42 and Catalog #: 321-45-46
  • Manufacturer

Manufacturer

  • 제조사 주소
    Exactech, Inc., 2320 NW 66th Ct, Gainesville FL 32653-1630
  • 제조사 모회사 (2017)
  • Source
    USFDA