table - Product Code JAI
원인
During exactrac robotics installation on the varian exact couch, the couch height position indication is re-calibrated to read out the correct positions. this re-calibration could cause the eventual failure of the vertical lift mechanism of the couch, which might result in patient injury or death.
조치
BrainLab issued a Product Notification Letter to the hospitals on 6/2/06 to advise the users not to lower the couch below 50 cm, and sent a second letter to the users on 6/7/06, informing them that BrainLab would remove the ExacTrac Robotics from the Exact Couch and re-calibrate the couch to restore the software limit switch. The Robotics will be re-installed once a permanent solution is available.
Nationwide, including the states of California, Colorado, Florida, Illinois, Maryland, Michigan, Missouri, New Hampshire, New York, North Carolina, Ohio, Pennsylvania, Tennessee and Texas
제품 설명
ExacTrac Robotics, sold alone as Robotic Tilt Module catalog 49720 and in a bundle as catalog 49700, for installation on the Varian Exact Couch, a component of the Novalis Shaped Beam Surgery System; a table tilt device - powered radiation therapy patient support assembly; BrainLab AG, Ammerthalstrasse 8, 85551 Heimstetten, Germany