Device Recall FastPlan Treatment Planning System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Varian Medical Systems, Inc. Oncology Systems 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    61193
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1217-2012
  • 사례 시작날짜
    2012-01-30
  • 사례 출판 날짜
    2012-03-16
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2013-10-21
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Accelerator, linear, medical - Product Code IYE
  • 원인
    An anomaly has been identified with the ct localization module of the fastplan treatment planning system. if the module reports a significant difference in tilt or spin on a slice and the use elects to "discard" that slice, an incorrect calculation of the isocenter location may result.
  • 조치
    "Urgent Medical Device Correction/Urgent Field Safety Notification" letters were sent on January 30, 2012 by certified mail to consignees in the US and by Federal Express for Canada. International sites will be notified by Varian International Customer Support Managers, who will translate letters if necessary. The notice provided a description of the issue, the actions that can be taken to avoid or mitigate the issue, and the steps Varian is taking to address the issue. Customer service number for US and Canada is: (888) 827-4265.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model HZ3, code numbers: HZ30001 HZ30556 HZ32037 HZ33064 HZ30044 HZ30588 HZ32045 HZ36303 HZ30110 HZ30624 HZ32050 HZ37011 HZ30019 HZ30557 HZ32039 HZ35070 HZ30054 HZ30594 HZ32046 HZ37002 HZ30111 HZ30630 HZ32051 HZ37015 HZ30021 HZ30571 HZ32040 HZ36101 HZ30061 HZ30604 HZ32047 HZ37003 HZ30112 HZ30640 HZ32052 HZ37027 HZ30032 HZ30572 HZ32042 HZ36102 HZ30077 HZ30612 HZ32048 HZ37007 HZ30131 HZ30646 HZ32053 HZ37101 HZ30033 HZ30573 HZ32044 HZ36302 HZ30108 HZ30617 HZ32049 HZ37010 HZ30144 HZ30649 HZ32054 HZ38006 HZ30154 HZ30656 HZ32055 HZ38062 HZ30258 HZ30743 HZ32071 HZ30403 HZ32013 HZ33020 HZ30510 HZ32028 HZ33047 HZ30167 HZ30670 HZ32056 HZ39001 HZ30265 HZ30762 HZ32072 HZ30404 HZ32014 HZ33021 HZ30511 HZ32029 HZ33049 HZ30171 HZ30675 HZ32058 HZ39002 HZ30273 HZ32001 HZ32073 HZ30407 HZ32017 HZ33022 HZ30513 HZ32032 HZ33050 HZ30178 HZ30681 HZ32059 HZ39004 HZ30280 HZ32002 HZ32074 HZ30426 HZ32018 HZ33026 HZ30520 HZ32033 HZ33053 HZ30181 HZ30684 HZ32060 HZ39007 HZ30283 HZ32003 HZ32075 HZ30460 HZ32019 HZ33030 HZ30531 HZ32034 HZ33055 HZ30182 HZ30703 HZ32061 HZ39010 HZ30293 HZ32005 HZ32076 HZ30461 HZ32020 HZ33031 HZ30537 HZ32035 HZ33059 HZ30190 HZ30712 HZ32062 HZ39011 HZ30294 HZ32006 HZ32077 HZ30462 HZ32021 HZ33035 HZ30540 HZ32036 HZ33062 HZ30207 HZ30727 HZ32063 HZ39012 HZ30312 HZ32007 HZ32084 HZ30485 HZ32022 HZ33039 HZ30208 HZ30730 HZ32064 HZ39014 HZ30330 HZ32008 HZ32086 HZ30495 HZ32024 HZ33041 HZ30224 HZ30731 HZ32065 HZ39015 HZ30349 HZ32009 HZ32108 HZ30500 HZ32025 HZ33042 HZ30230 HZ30737 HZ32067 HZ39016 HZ30368 HZ32010 HZ32125 HZ30508 HZ32026 HZ33044 HZ30251 HZ30740 HZ32069 HZ39997 HZ30398 HZ32012 HZ32126 HZ30509 HZ32027 HZ33046.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - Nationwide (USA) and the countries of Canada, China, Germany, Ireland, Spain, Singapore, Georgia, Hong Kong, India, Macau, Malaysia, Russia, Sri Lanka, Taiwan, Japan, Chile, Columbia, Venezuela and Australia.
  • 제품 설명
    FastPlan Treatment Planning System, Model HZ3, Varian Medical Systems, Palo Alto, CA. Varian FastPlan Treatment Planning System is a radiation treatment planning system composed of a high resolution UNIX graphics computer, a printer, a digitizing tablet, and radiation treatment planning application software.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Varian Medical Systems, Inc. Oncology Systems, 911 Hansen Way, Palo Alto CA 94304-1028
  • 제조사 모회사 (2017)
  • Source
    USFDA