Device Recall Fiber stripper 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 American Medical Systems Innovation Center - Silicon Valley 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    71534
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2175-2015
  • 사례 시작날짜
    2015-05-28
  • 사례 출판 날짜
    2015-07-21
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2016-01-19
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Powered laser surgical instrument - Product Code GEX
  • 원인
    Validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (ifu) for the sureflex reusable fibers and the reusable stripper and cleaver accessories do not meet the new guidance on qualifying reusable products.
  • 조치
    Urgent Medical Device Field notification letters were sent on June 9, 2015. Translation for countries out of the US will be conducted. Communications are being coordinated by the Prostate Health QA Director. Urgent Medical Device Field notification Letters advised customers of the issue and list all affected products with name and model numbers. Any SureFlex Reuseable Fibers which have not been used but are in inventory and with the 3 year shelf life should be returned as per instructions enclosed. Any of the SureFlex devices in inventory that have been used should be discarded or returned. All Reuseable Stripper and Cleaver Accessories that have not been used should be returned. Used devices should be destroyed or returned. Questions to 1-408-456-3320. The Acknowledgement form should be completed and returned.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model 80-10002-002; All product manufactured from April 8, 2012 to April 8, 2015
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
  • 제품 설명
    Tool, Stripper W/BLD, 12S, Autoclavable; || The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.