Device Recall FreeStyle Flash Blood Glucose Monitors 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Abbott Diabetes Care, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    67472
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-1211-2014
  • 사례 시작날짜
    2014-02-19
  • 사례 출판 날짜
    2014-03-18
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2015-08-11
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    System, test, blood glucose, over the counter - Product Code NBW
  • 원인
    Abbott diabetes care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with free style test strips lots within expiry.
  • 조치
    The firm, Abbott Diabetes Care, sent an "IMPORTANT: URGENT PRODUCT RECALL"- letter dated February 14, 2014 to all customers for whom they have a record of receiving 0mV meters on February 19, 2014 via trackable method. Abbott also issued Press Release on March 14, 2014. The letter describe the product, problem and actions to be taken. The customers were instructed to discontinue use of your FreeStyle Flash or FreeStyle Blood Glucose Meter and call Abbott Diabetes Care Customer Service at 1-888-345-5364 for replacement at no charge; upon receiving replacement meter; immediately dispose of your old meter in accordance with your local and state disposal regulations; and please read your test strip insert carefully to ensure that you are using the correct test strip type for your Blood Glucose Meter. The firm, OmniPod, sent a ACD letter dated February 7, 2014 and included Abbotts February 14, 2014 letter to its customers. The letter described the product, problem, and actions to be taken. The customers were informed that you can safely continue to use the OmniPod System for your insulin delivery needs. The customers were instructed to read all of the materials included in the Abbott communication carefully and make sure you understand how it affects you; to immediately discontinue use of FreeStyle test strips with the Freestyle meter built into the OmniPod system; call Abbott Diabetes Care Customer Service immediately at 1-877-584-5159 for replacement strips; while waiting on replacement strips, use an unaffected meter such as the FreeStyle Freedom blood glucose meter; you can always use an alternative blood glucose meter. The notification will be posted on the Abbott Diabetes Care website at 222.abbottdiabetescare.com. If you have any further questions, please call Abbott Diabetes Care Customer Service at 1-877-584-5159. For questions or concerns with the continued use of your OmniPod Insulin Management System contact Insulet Product Support at 1-800-591-34

Device

  • 모델명 / 제조번호(시리얼번호)
    Part number CAT11001-0; all serial numbers
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.
  • 제품 설명
    FreeStyle Blood Glucose Monitors System. || For in vitro diagnostic use only. Store at room || temperature below 86 degrees F (30 degrees C). Use test strips || only within the system operating temperature || range as outlined in your Owner's Booklet. || Meter made in China; || Distributed by Abbott Diabetes Care, Alameda, CA || The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete can product inaccurate results. || The FreeStyle Glucose meter is incorporated into the OmniPod insulin Management system (manufactured by Insulet Corporation, Bedford, MA and is intended for subcutaneous delivery of insulin. || Intended to monitor blood glucose from samples taken from the body. IVD use only,
  • Manufacturer

Manufacturer

  • 제조사 주소
    Abbott Diabetes Care, Inc., 1360 S Loop Rd, Alameda CA 94502-7000
  • Source
    USFDA