Device Recall G1 High Speed Electric Handpiece Air Cooled (eG1A) 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 The Anspach Effort, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    74447
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2253-2016
  • 사례 시작날짜
    2016-06-15
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2016-12-14
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Motor, drill, electric - Product Code HBC
  • 원인
    The graphics at the attachment interface indicating locked and unlocked position are reversed.
  • 조치
    The firm, DePuy Synthes, sent an "URGENT NOTICE: MEDICAL DEVICE-RECALL (Removal)" letter dated June 15, 2016 to its customers. The letter described the product, problem and actions to taken. The customers were instructed to review your inventory and immediately remove the product subject to this recall from stock; If you DO HAVE any of the identified devices, please take the following steps: " Ensure anyone in your facility impacted by this notification reads this letter carefully. Customer Letter Complete the Verification Section (page 3 of this letter) by checking the appropriate box indicating product subject to the recall has been located. Indicate the number of devices found. " The RMA information has been filled in for your convenience. " Please include your name, title, address, telephone number, signature and date in the spaces provided. " Send a copy of the completed Verification Section to DePuy Synthes, Customer Quality Department by: " Fax: (561) 627-2682 " Scan/email: DPYUS-PowerToolsFieldActions@its.jnj.com " Giving it to your DePuy Synthes Sales Consultant. " This return documentation acknowledges your receipt of medical device removal information. " Return the affected product to your DePuy Synthes Sales Consultant " Keep this notice visibly posted for awareness until all products subject to this recall has been returned. While processing your returns, please maintain a copy of this notice with the product subject to this action and keep a copy for your records. If you DO NOT HAVE the identified product, please take the following steps: " Complete the attached Verification Section (page 3 of this letter) by checking the appropriate box indicating that no affected product has been located. Include your name, title, address, telephone number, signature and date in the spaces provided. " Send a copy of the completed Verification Section to DePuy Synthes, Customer Quality Department by Fax: (561) 627-2682 " Scan/email: DPYUS-PowerToolsFiel

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot #'s: H29309439804, H44310033104, K09311396703, and K09311396704.
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution: US Distribution to states of: GA and FL; and country of: Switzerland.
  • 제품 설명
    Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) || Use for cutting and shaping bone including bones of the cranium and spine.
  • Manufacturer

Manufacturer

  • 제조사 주소
    The Anspach Effort, Inc., 4500 Riverside Dr, Palm Beach Gardens FL 33410-4235
  • 제조사 모회사 (2017)
  • Source
    USFDA