Digital Radiographic System - Product Code KPR
원인
Loss of image data: software anomaly in the processing software on the definium 8000 system may impact patient safety when using the volumerad advanced application (option). the slice visualization of volumerad exams acquired on the wallstand receptor will be offset by 17mm. this prevents the visualization of anatomy within a distance of 17mm from the wallstand patient barrier (receptor cover).
조치
GE sent a Product Safety Notification, dated March 7, 2008, to all customers. The letter described the safety issue, products affected, and stated that a GE Healthcare Field Engineer will schedule a visit to each facility to install a software modification.