X-Ray System - Product Code KPR
원인
The actual average illuminance for the collimators are approximately 140-lux. this does not meet the 160-lux requirement. ge will be correcting the units by replacing all collimators.
조치
GE will be correcting the units by replacing all collimators that don't meet the 160-lux requirement. The field correction will be initiated and is scheduled to be completed by October 2008.
Worldwide Distribution --- USA including states of CA, DE, FL, GA, ID, IL, KS, KY, MD, MI, MO, MT, NC, NJ, NM, NY, OH, PA, SC, TX, UT, VA, and WV, and countries of Australia, Canada, China, France, Germany, Greece, Italy, Korea, Poland, Saudi Arabia, United Kingdom, and Venezuela.
제품 설명
GE Healthcare Proteus XR/a Eclipse Collimator model # 2379827.