Device Recall Genetic Screening Processor (GSP 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 PerkinElmer Life and Analytical Sciences, Wallac, OY 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    76165
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1107-2017
  • 사례 시작날짜
    2017-01-05
  • 사례 출판 날짜
    2017-01-27
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-08-24
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Fluorometer, for clinical use - Product Code KHO
  • 원인
    The firm became aware that the ball bushing axels are defective in the gsp instrument. during exposure to humid conditions, there is a risk that the axle material will corrode. deterioration of the axle can be observed as deposits of corrosion on the defective axles. ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.
  • 조치
    The recall was initiated on January 5, 2017 via email sent through the PerkinElmer Technical Support group located in Akron Ohio. Actions to be taken by user: 1. As instructed in the analyte specific kit inserts, the use of Kit Controls is required to assure the day-to-day validity of results. Patient results should only be reported if control results for the assay meet the laboratorys established criteria for acceptability. 2. If a movement error occurs customers will be directed to contact their local PerkinElmer field service representative. The movement error may indicate deterioration of the axle and therefore the use of the instrument should be avoided until field service personnel have inspected the status of the ball bushing axles or replaced affected ball bushing axles. Actions to be taken by the PerkinElmer: As a corrective action, PerkinElmer will replace the affected ball bushing axles in all the GSP instruments identified in this FSCA/Recall. The replacement may take place either as part of the initial installation activity before the instrument is used, scheduled periodic maintenance, or during a dedicated service visit. This field correction is considered as mandatory by the manufacturer. Customers will be contacted by PerkinElmer to arrange the field service visit.

Device

  • 모델명 / 제조번호(시리얼번호)
    Serial Numbers: 20210239 20210243 20210244 20210245 20210246 20210247 20210248 20210249 20210250 20210251 20210252 20210253 20210254 20210255 20210256 20210257 20210258 20210259 20210260 20210261 20210262 20210263 20210264 20210265
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    MA, SC, TX, AL OUS: China, Mexico, Peru, Morocco, Italy, Finland, Ecuador, Brazil
  • 제품 설명
    Genetic Screening Processor (GSP); PerkinElmer Part Number= 2021-0010 || Intended for in vitro quantitative and qualitative determination of analytes in body fluids.
  • Manufacturer

Manufacturer

  • 제조사 주소
    PerkinElmer Life and Analytical Sciences, Wallac, OY, Mustionkatu 6, P.O. Box 10, Turku Finland
  • 제조사 모회사 (2017)
  • Source
    USFDA