Device Recall GlideScope Titanium SingleUse Video Laryngoscope 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Verathon, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    73014
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-1035-2016
  • 사례 시작날짜
    2016-01-29
  • 사례 출판 날짜
    2016-03-11
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2017-03-01
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Laryngoscope, rigid - Product Code CCW
  • 원인
    The firm has become aware of the potential for disruption (flickering) in the video laryngoscopy image when glidescope titanium su laryngoscope blades are in use. video "flickering" appears as the intermittent break-up of the on-screen video image, appearing as either distorted horizontal or vertical bars of displaced video signal.
  • 조치
    The firm, Verathon, began mailing the "URGENT: MEDICAL DEVICE RECALL" GLIDESCOPE VIDEO LARYNGOSCOPE TITANIUM SINGLE-USE BLADES letter to their consignees on January 29, 2016. The letter describes the product, problem and actions to be taken. Verathon will instruct the company's foreign distributors to communicate the Customer Recall Letter to their international customers. Verathon has contracted with a business services company, Stericycle, to coordinate the initial consignee communications in the United States, Australia and Canada. Stericycle will send the Customer Recall Letters, and conduct consignee follow-up communications for return of the Recall Response Forms. Verathon will replace all affected GlideScope Titanium Single Use (SU) blades that remain in your facility's inventory with new product that has been passed the enhanced screening test for video flickering. Customers were advised to take the following actions: -Fill out the attached Recall Response Form and return it to Verathon by fax (888) 943-5170, or email, verathon5165@stericycle.com. Please return the form even if you do not have any blades subject to the recall. If consignees need to expedite a product replacement contact Verathon representative or Verathon Customer Care at (800) 331-2313 (U.S. and Canada). The firm has live agent support Monday through Friday from 8:00am to 5:00pm Eastern Time. Consignees can also send email to CCTechSupport@verathon.com. Consignees who have any questions about this Product Recall, please contact Verathon representative or call (888) 943-4207 (U.S. and Canada) or send email to verathon5165@stericycle.com. Please report any suspected malfunction or adverse event related to any GlideScope Video Laryngoscope devices to Verathon Customer Care at the telephone numbers above and to FDA's MedWatch Program (1-800-FDA-1088 or www.fda.gov/medwatch).

Device

  • 모델명 / 제조번호(시리얼번호)
    LOT Number ranges:  -LoPro S3: 081814 - 093015; -LoPro S4: 081114 - 090315; -MAC S3: 080814 - 101315; -MAC S4: 022514 - 082115;
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US (nationwide) and to countries of: Australia, Belgium, Canada, Denmark, France, Germany, Hong Kong, Italy, Luxembourg, Saudi Arabia, Singapore, Spain, Sweden, The Netherlands, U.A.E., and United Kingdom.
  • 제품 설명
    GlideScope Titanium Single-Use Video Laryngoscope. It is a single-use video laryngoscope system (GlideScope Titanium System) consists of a sterile-packaged, single-use video laryngoscope, multi-use Smart Cable and portable GlideScope video monitor. || The system incorporates a high-resolution, full-color digital camera and monitor for real-time viewing and recording. || Affected Product Names and Part Numbers: || -LoPro S3: 0574-0130 (single blade) or 0270-0769 (box of 10) || -LoPro S4: 0574-0131 (single blade) or 0270-0770 (box of 10) || -MAC S3: 0574-0132 (single blade) or 0270-0771 (box of 10) || -MAC S4: 0574-0133 (single blade) or 0270-0772 (box of 10) || The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Verathon, Inc., 20001 N Creek Pkwy, Bothell WA 98011-8218
  • 제조사 모회사 (2017)
  • Source