Device Recall Graphic Case 6.5mm & 7.3mm Cannulated Screw Sets 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Synthes (USA) Products LLC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79754
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-2245-2018
  • 사례 시작날짜
    2018-03-05
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Tray, surgical, instrument - Product Code FSM
  • 원인
    Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.
  • 조치
    On March 8, 2018, the recalling firm sent Urgent Product Recall letters to its customers to inform them of the recall. Customers were advised to take the following actions: 1. Immediately review their inventory to identify and quarantine all affected products listed above in a manner that ensures the affected products will not be used. 2. Complete the Verification Section and send a copy to the recalling firm. 3. Forward this notice to anyone in the facility that needs to be informed. 4. If any of the affected products have been forwarded to another facility, contact that facility to arrange return. Customers with questions were advised to call 610-719-5450, 8AM-5PM (EST), Monday through Friday, or contact their DePuy Synthes sales consultant.

Device

  • 모델명 / 제조번호(시리얼번호)
    FL00860
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.
  • 제품 설명
    Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set
  • Manufacturer

Manufacturer

  • 제조사 주소
    Synthes (USA) Products LLC, 1301 Goshen Pkwy, West Chester PA 19380-5986
  • 제조사 모회사 (2017)
  • Source
    USFDA