Tonometer (Calibration And Q.C. Of Blood-Gas Instruments), Clinical - Product Code LCH
원인
The ethanol standard is oos on the high end. the results were 1.58 mg/ml while the specification is 1.477-1.538 mg/ml.
조치
All customers were nofitied by phone and letter. The first notification occurred on 12/17/2002 followed with additional notifications on 1/24/2003, 1/28/2003, and 2/4/2003.
The Ethanol Products were distributed nationwide by the following distributors: VWR International, 1310 Goshen Pkwy, West Chester, PA 19380; Government Scientific, 8460 K-Tyco Road, Vienna, VA 22182; Allegiance Healthcare, 1430 Waukegan Rd., McGaw Park, IL 60085; Fisher Scientific, 2000 Park Lane, Pittsburgh, PA 15275; Hawaii Chemical Co., 2363 North King Street, Honolulu, HI 96819; Krackeler Scientific Inc., 57 Broadway, Albany, NY 12202; Midland Scientific Inc., 1202 South 11th St., Omaha, NE 68108; Reagents, Inc. 4746 Sweden Rd., Charlotte, NC 28224. The Papanicolaou Stain was distributed nationwide by the following distributors: VWR International, 1310 Goshen Pkwy, West Chester, PA 19380; Fisher Scientific, 2000 Park Lane, Pittsburgh, PA 15275; Kramer Scientific, 711 Executive Blvd, Valley Cottage, NY 10989; Laboratory Supply Company, 3069 Mercury Rd., Jacksonville, FL 32207; Allegiance Healthcare, 1430 Waukegan Rd., McGaw Park, IL 60085; Infolab, 3000 Greenhill Dr., Round Rock, TX 78664; Government Scientific, 8460 K-Tyco Road, Vienna, VA 22182.
제품 설명
Ethanol Standard, 1.5mg Ethanol in 1 mL. Product size configuration: 1 mL glass ampule shipped 10 ampules per case.