Device Recall HeartMate II Left Ventricular Assist System 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Thoratec Corporation 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    61268
  • 사례 위험등급
    Class 1
  • 사례 연번
    Z-1228-2012
  • 사례 시작날짜
    2012-02-23
  • 사례 출판 날짜
    2012-03-23
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2012-11-08
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Ventricular (assisst) bypass - Product Code DSQ
  • 원인
    A trend of disconnected bend reliefs on the sealed outflow graft used with the heartmate ii lvas has been identified.
  • 조치
    Thoractec Corporation sent an Urgent Medical Device Correction letter on February 23, 2012 via overnight courier and e-mail to all affected consignees. The Urgent Medical Device Correction notification letters identify the affected products, description of the problem with illustrations, symptoms of problem and immediate actions to taken. All users are asked to review the attached labeling, including new caution statements and share with all surgical personnel. Consignees are instructed to completed and returned the Acknowledgement Form indicating that the information was received and understood. Additionally, an update concerning the Worldwide Medical Device Correction Notification is enclosed with each letter. Questions should be directed to Thoratec Regulatory Affairs at 925-847-8571.

Device

  • 모델명 / 제조번호(시리얼번호)
    Sealed Outflow Graft: Catalog number 103393, all lot numbers; Sealed Outflow Short Bend Relief; Catalog number 104692, all lot numbers; Catalog number 104911, HM II LVAS Implant Kit (with sealed grafts) [US labeling] (contains catalog number 103393); all serial numbers; Catalog number 104912, HM II LVAS Implant Kit (with sealed grafts) [CE Mark labeling], contains catalog no. 103393); all serial numbers:
  • 의료기기 분류등급
  • 의료기기 등급
    3
  • 이식된 장치?
    Yes
  • 유통
    Worldwide Distribution - USA (nationwide) and the countries of: Austria, Belgium, Denmark, Germany, Greece, France, Hong Kong, Kuwait, Ireland, Israel, Italy, Lebanon, Liechtenstein, Lithuania, Luxembourg, Malaysia, Netherlands, Turkey, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Thailand and United Kingdom.
  • 제품 설명
    HeartMate II Left Ventricular Assist System (HM II LVAS) Sealed Outflow Graft Bend Relief. || The Sealed Outflow Graft Bend Relief is provided as a component of the HM II Sealed Outflow Graft, which in turn is a component of the HM II LVAS Implant Kit with Sealed Grafts. A short configuration of the HM II LVAS Sealed Outflow Graft Bend Relief is provided as a stand-along product. || Thoratec Corporation || 6035 Stoneridge Drive || Pleasanton, CA 94588. || Product Usage: || The HM II LVAS is intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; either for temporary support, such as a bridge to Cardiac Transplanation or myocardial recovery, or as permanent destination therapy. The HeartMate II is intended for use inside or outside the hospital. The outflow of the HM II LVAS blood pump is connected to the aorta with a flexible vascular graft. The outflow graft is provided with a bend relief, a three or four in long tube of reinforced ePTFE surrounding the outflow graft proximal to the pump that is designed to prevent kinking of the outflow graft.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Thoratec Corporation, 6035 Stoneridge Dr, Pleasanton CA 94588-3270
  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    USFDA