Device Recall Hospira LifeCare PCA Label Utility, Version 1.0 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Hospira Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    46846
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1181-2008
  • 사례 시작날짜
    2008-01-03
  • 사례 출판 날짜
    2008-08-06
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2010-03-09
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • 데이터 추가 비고
    pharmacy compounding software - Product Code MEA
  • 원인
    Incorrect dosage labels created: customers have experienced independent changes in drug concentration (mg/ml to mcg/ml and vice versa) and expiration dates, and the addition of trailing zeros to the barcodes.
  • 조치
    Hospira sent letters dated 1/3/08 to the immediate recipients of the product via FedEx overnight letter on the same date. The letters informed the accounts of the reports of independent changes in drug concentration while editing a saved formulation using the Hospira LifeCare PCA Label Utility Software CD that was provided to some customers in a Hospira MedNet Drug Library Creation Team Introductory Kit-LifeCare PCA kit. The accounts were requested to remove the utility from any PC in their facility on which it is currently installed, destroy the CD and sign and return the enclosed removal verification form verifying that they have removed the software from their system. The letter listed commercially available software packages that the account could validate and implement in their facility. Any questions concerning the utility software withdrawal were directed to Hospira at 1-800-241-4002, option 4, then option 1.

Device

  • 모델명 / 제조번호(시리얼번호)
    Version 1.0, all CDs
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Nationwide Distribution and Canada
  • 제품 설명
    Hospira LifeCare PCA Label Utility, Version 1.0 CD. The installation/user guide is entitled "LifeCare PCA Barcode Label Utility Program." This software program is designed to allow entry of each drug concentration into a formulation library on a pharmacy computer for standardized printing of barcode labels for compounded drug product vials.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Hospira Inc., 275 N Field Dr, Lake Forest IL 60045-2579
  • 제조사 모회사 (2017)
  • Source
    USFDA