Device Recall IMRIS ORT100 Table 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Deerfield Imaging, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79360
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0975-2018
  • 사례 시작날짜
    2018-02-20
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Table, radiologic - Product Code KXJ
  • 원인
    The rotational locking mechanisms on the ort 100/ t2x tables can become disengaged or unlocked while in use which could potentially result in unintended movement of the table.
  • 조치
    The firm disseminated the notices on 02/23/2018 and followed with emails on 02/26/2018. The notices stated: "To minimize the risk of injury, IMRIS is taking the following corrective actions: " ORT100 Table: The rotational locking mechanism shall be replaced in all ORT100 tables that do not contain the most current design. " T2X Table: Tables will have their rotational lock inspected. If components are worn or performance is other than intended, the rotational lock assembly will be serviced or replaced entirely. IMRIS has identified that your facility has one or more of these OR Tables which require evaluation. Please confirm details on page 2 and return your response to IMRIS.IMRIS Customer Service will contact you to implement this corrective action. IMRIS recommends these preventive actions: " Per the ORT Operator Manuals the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, back section, headrest, pendant connection and swivel release pedal free engagement " During usage, verify the table is locked by manually pushing the table to make sure it does not move rotationally. " If the rotational lock is observed to be not functioning properly, please contact IMRIS Customer Service. The appropriate competent authorities have been notified of this corrective action. IMRIS is committed to patient and user safety, and apologizes for any inconvenience that this may cause. Please contact customer support via our service line at 1-866-475-0525 or Email: customersupport@i mris.com if you have any questions or concerns."

Device

  • 모델명 / 제조번호(시리얼번호)
    Serial Numbers: 10002341, 10003629, 10003801, 10002654, 10002795, 10002796, 10002847, 10003830, 10003243, 10002848, 10003241, 10005651, 10007348, 10007424, 10006886
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    US, Canada, Australia, China, Qatar
  • 제품 설명
    IMRIS ORT100 Table, Part Number 110470-000