Device Recall IMRIS ORT300 Removable Operating Room Table 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Deerfield Imaging, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    79366
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1218-2018
  • 사례 시작날짜
    2018-02-23
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Table, operating-room, ac-powered - Product Code FQO
  • 원인
    The rotational locking mechanisms on the ort 200/300 tables fail to operate such that it cannot change states between locked and unlocked.
  • 조치
    The firm disseminated their notices on 02/23/2018 by mail and followed with an email on 02/26/2018. The notice stated: "It has recently come to our attention that the ORT200/300 rotational lock fails to operate such that it cannot change states between locked and unlocked. This may pose a risk of injury to the patient. To minimize the risk of injury, IMRJS is taking the following corrective actions: " ORT200 and 300 tables: all customer sites with these tables shall have the rotational lock operation verified and adjusted, if necessary. IMRIS has identified that your facility has one or more of these OR Tables which require evaluation. Please confirm details on page 2 and return your response to IMRIS. IMRIS Customer Service will contact you to implement this corrective action. IMRIS recommends these preventive actions: " Per the OAT Operator Manuals the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, tabletops, headrest, pendant connection and rotation. " During usage, verify the table is locked by manually pushing the table to make sure it does not move rotationally. " If the rotational lock is observed to be not functioning properly, please contact IMRIS Customer Service. The appropriate competent authorities have been notified of this corrective action. IMRIS is committed to patient and user safety, and apologizes for any inconvenience that this may cause. Please contact customer support via our service line at 1-866-475-0525 or Email: customersupport@imris. com if you have any questions or concerns."

Device

  • 모델명 / 제조번호(시리얼번호)
    Serial numbers: 10005695, 10002998, 10005456, 10007411, 10007377, 10002997, 10002996, 10005612, 10004372, 10002841, 10004515, 10005613, 10002255
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    US, Canada, China, Sweden, Germany, Japan, Qatar, France, Audtralia
  • 제품 설명
    IMRIS ORT300 Removable Operating Room Table, 50-60 Hz Version is intended for use during diagnostic examinations or surgical procedures to support and position a patient.
  • Manufacturer

Manufacturer